Total hemispherical emittance measurement apparatus for solar selective coatings

by Albuquerque Sandia Laboratories, NM

Technical Report, 1975

Barcode

CSP Unique ID 190709054

Status

Electronic Resource

Call number

**Click on MARC view for more information on this report.**

Publication

SAND Report: SAND75-0079, June 1975.

Language

Library's review

ABSTRACT:
An apparatus designed to measure the total hemispherical emittance of solar selective coatings and metallic substrates over the temperature range o0 c to approximately 300°c is described. This apparatus utilizes a delta-power calorimetric measurement technique. After an error analysis is
Show More
presented, emittance values of several metallic substrate material are presented and compared with previous measurements.
Show Less
Page: 0.0731 seconds